7164, 7166,7167, and 7187 - A Gate Signal is applied to
the front panel which enables all channels to accept data.
Each channel has a "Sample/Hold" circuit. During the Gated
time, the module is in "Sample" mode. (The Start pulse in the 7187
begins the "Sample" mode). Once the Gate Signal turns off, the channels
go into "Hold" mode (The Stop pulse in the 7187 switches to "Hold" mode.)
locking out the front end and keeping the module from accepting
any more data until cleared. The CONVERSION TIME is the time it takes
to multiplex the data from all 16 "Sample/Hold" circuits through
the ADC and into memory It is 7.2us and timing begins at the back edge
of the GATE pulse.
7186 - The Start pulse begins "Sample" mode and
the Stop pulse switches to "Hold" mode. This module uses timing
from the COMMON signal to start the Conversion Cycle. In COMMON START
mode, some ranges require additional conversion delay as described
below*. The CONVERSION TIME is 7.2us and the timing begins at
the leading edge of the COMMON pulse, although the actual conversion begin
750ns later.
In ALL these modules, digitization may be delayed by an
amount from 0 to 16us in 62.5ns increments using the jumpers at
the rear of the module. This will allow a greater acceptance window
for CLEAR signals if needed. Once digitization begins, you will not
be able to clear or access the module until the conversion cycle has completed.
Conversion Time for each Module is 7.2us, which
includes all 16 Channels
| ADC 7164 |
Conversion starts 600ns
after the trailing edge of the Gate signal. |
| QDC 7166, 7167 |
Conversion starts 750ns
after the trailing edge of the Gate signal |
| TDC 7186, 7187 |
Conversion starts 750ns
after the leading edge of the COMMON pulse for the 7186, and 750ns after the trailing edge of the Gate pulse for the 7187 |
|
The 600ns and 750ns delay before starting the Conversions is
for settling time and also allows time to accept fast clear signals. |
|
*7186 COMMON START ONLY
It is recommended to delay digitization for the 7186
TDC when using COMMON START mode greater than 100ns full scale. This
prevents the module from beginning the Conversion cycle to soon and
still allows settling time and a few hundred nanoseconds acceptance window.
Example: If the module is set for 2us Full Scale
with no additional delay, after 750ns from the Common Pulse (START),
digitization begins. The module switches to "Hold" mode before
Full Scale is reached therefore the highest reading you could get
is around 750ns instead of 2us, This would also happen in channels
with no stop pulses. The Manual shows the minimum recommended delays
for the various Full Scale ranges when used in COMMON START mode. The
chart is repeated here. Note: The COMMON STOP mode does not require
this.
| RANGE |
DELAY |
JUMPER POSITION |
| 100 nsec |
0 |
- |
| 200 nsec |
125 nsec |
2 |
| 400 nsec |
312.5 nsec |
4,1 |
| 800 nsec |
750 nsec |
8,4 |
| 1 usec |
937.5 nsec |
8,4,2,1 |
| 2 usec |
1937.5 nsec |
16,8,4,2,1 |
| 4 usec |
3937.5 nsec |
32,16,8,4,2,1 |
| 8 usec |
7937.5 nsec |
64,32,16,8,4,2,1 |
| Other ranges |
> (Range
- 100 nsec) |
Back to Technical Questions